锥形束CT测量牙槽骨高度准确性和可靠性研究
其他题名Accuracy and reliability of cone-beam computed tomography in measuring alveolar bone height
冯驭驰1; 许天民1
关键词锥形束ct 牙槽骨高度 Cbct Alveolar Bone Height
刊名中华口腔正畸学杂志
2012
DOI10.3760/cma.j.issn.1674-5760.2012.03.003
19期:3页:136-140
收录类别中国科技核心期刊
文章类型Journal Article
摘要目的 评价在锥形束CT(cone beam compnted tomography,CBCT)图像上测量牙槽骨高度的准确性和可靠性.方法 8个人类湿颅标本经CBCT扫描后测量图像上牙尖点到釉牙骨质界距离(T-CEJ)、牙尖点到牙槽嵴顶距离(T-BM)、釉牙骨质界到牙槽嵴顶距离(CEJ-BM).在标本上实际测量以上项目.将两种测量值进行配对t检验.选择一定数量的位点重复测量,计算组内相关系数(ICC),进行一致性检验.结果 CBCT测量值均略小于标本实际测量值.其中对T-CEJ的测量两种方法无统计学差异,差值均值为-0.041 mm;对T-BM和CEJ-BM的测量两种方法有统计学差异,差值均值分别为-o.268 mm和-0.208 mm.所有重复测量的ICC均大于0.96.结论 CBCT测量牙槽骨高度与标本测量虽然存在统计学差异,但差异较小,在临床实际可接受范围之内.CBCT测量的可靠性较好. Objective To evaluate the accuracy and reliability of cone-beam computed tomography (CBCT) in measuring alveolar bone height.Methods Eight human cadaver head samples were scanned with CBCT technology.Measurements were made on each tooth from the cusp or the middle point of incisal margin to the cementoenamel junction (T-CEJ),from the cusp or the middle point of incisal margin to the alveolar bone margin (T-BM) and from the cementoenamel junction to the alveolar bone margin (CEJ-BM) along the long axis of the tooth.Measurements were made both in the CBCT images and on the real samples.The paired-t test was made between the two different measurements to evaluate the accuracy of the CBCT measurements.And 210 sites from 6 samples were measured.The Interclass correlation coefficient (ICC) was calculated to evaluate the reliability of CBCT measurements.Results The values of CBCT measurement were smaller than those of the direct measurement.Significant difference was found in the values of T-BM and CEJ-BM.The mean deviations were - 0.268 mm and - 0.208 mm.The values of T-CEJ showed no significant difference and the mean deviation was -0.041 mm. The ICC of all the measurement was above 0.96.Conclusions Significant difference was found in the measurement of alveolar bone height between CBCT and direct measurement. But the mean deviation was very small. The CBCT measurement showed high reliability.
语种中文
原文出处查看原文
引用统计
文献类型期刊论文
条目标识符http://ir.bjmu.edu.cn/handle/400002259/42360
专题北京大学第一临床医学院_口腔科
作者单位1.北京大学第一医院口腔科
2.100081,北京大学口腔医学院·
3.口腔医院正畸科
推荐引用方式
GB/T 7714
冯驭驰,许天民. 锥形束CT测量牙槽骨高度准确性和可靠性研究[J]. 中华口腔正畸学杂志,2012,19(3):136-140.
APA 冯驭驰,&许天民.(2012).锥形束CT测量牙槽骨高度准确性和可靠性研究.中华口腔正畸学杂志,19(3),136-140.
MLA 冯驭驰,et al."锥形束CT测量牙槽骨高度准确性和可靠性研究".中华口腔正畸学杂志 19.3(2012):136-140.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[冯驭驰]的文章
[许天民]的文章
百度学术
百度学术中相似的文章
[冯驭驰]的文章
[许天民]的文章
必应学术
必应学术中相似的文章
[冯驭驰]的文章
[许天民]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。