100 kVp条件下碘克沙醇(270 mg I/mL)应用于头颈部CTA 检查的可行性研究
其他题名FeasibilityofCTangiographyofheadandneckusing100kVpastubevoltagewithiodixanol(270mgI/mL)ascontrastmedium
赵永为; 郭小超; 王霄英; 雷平贵; 刘建新; 张保翠; 高福生
关键词脑动脉 颈动脉 血管成像 体层摄影术 对比剂 X线计算机 辐射量 体型特异性剂量估计 Cerebral Artery Carotid Artery Ct Angiography Contrast Medium Radiation Dose Size-specific Dose Estimate
刊名放射学实践
2014
4页:365-368
收录类别中国科技核心期刊
文章类型Journal Article
摘要目的:探讨100 kVp条件下应用碘克沙醇(270 mg I/mL)行头颈部CTA检查的可行性。方法:连续40例行头颈部CTA检查的患者,根据检查方案分为A、B两组进行评价。A组:使用碘帕醇(370 mg I/mL),管电压120 kVp,噪声指数(NI)13,共20例;B组:使用碘克沙醇(270 mg I/mL),管电压100 kVp,NI 11,共20例。读取CT剂量报告中的容积剂量指数(CTDIvol),计算体型特异性剂量估计(SSDE)值。图像质量主观评价:采用双盲法由两位具有3年以上CT诊断经验的医师采用5分法进行评价。图像质量客观评价:测量并分析两组中主动脉、颈总动脉、颈内动脉及大脑中动脉的CT值和标准差(SD)。结果:A、B组中患者年龄[分别为(64.55±8.76)岁和(63.65±14.06)岁]和BMI[(23.92±2.23)和(23.56±2.20)kg/m2]的差异无统计学意义(P>0.05)。A、B 两组间 CTDIvol[分别为(33.39±2.53)和(26.76±1.03)mGy]和SSDE[分别为(66.27±4.01)和(53.58±2.06)mGy]的差异有统计学意义(P<0.05)。A、B组间主观评分(分别为4.20±0.62和4.10±0.55)的差异无统计学意义(P>0.05)。客观评价:两组间主动脉、颈总动脉、颈内动脉和大脑中动脉的CT值及SD值的差异均无统计学意义(P>0.05)。结论:100 kVp条件下、应用碘克沙醇(270 mg I/mL)行头颈部CTA检查,在图像质量保持不变的情况下,可降低辐射剂量并减少对比剂使用量。 Objective:To evaluate the feasibility of using 100kVp as tube voltage and 270mg I/mL iodixanol as con-trast medium in head and neck CT angiography (CTA ).Methods:Forty consecutive patients underwent head and neck CTA.Patients were divided into group A and group B according to different CT protocols.Group A:20 patients,Iopamidol (370mg I/mL)as contrast medium,tube voltage 120kVp,noise Index 13;group B:20 patients,Iodixanol (270mg I/mL)as contrast medium,tube voltage 100kVp,noise Index 11.CT dose index (CTDIvol)was recorded and size-specific dose esti-mate (SSDE)was calculated.The subjective evaluation of CTA image quality was obtained by 2 radiologists having more than three years experience of CT diagnosis,using double blind method with five scale score.CT attenuation and standard deviation of vessel lumen (including aorta,common carotid artery,internal carotid artery)were measured and used as objec-tive evaluation of image quality.Results:There was no significant statistic difference in age [(64.55 ± 8.76 )years vs (63.65±14.06)years,P>0.05]and BMI [(23.92±2.23)kg/m2 vs (23.56±2.20)kg/m2 ,P>0.05]of two groups.The mean values of CTDIvol [(33.39±2.53)mGy vs (26.76±1.03)mGy,P<0.05]and SSDE [(66.27±4.01)mGy vs (53.58±2.06)mGy,P<0.05]in Group A were higher than those of group B,with significant statistic difference.There was neither significant difference in subjectively evaluated image quality (4.2±0.62 vs 4.1±0.55,P>0.05),nor in objec-tively evaluated image quality showing as CT values of blood vessel lumen between the 2 groups,as well as the standard de-viation of CT value (P>0.05).Conclusion:Head and neck CTA with tube voltage as 100kVp and Iodixanol (270mg I/mL) as contrast medium can reduce radiation dose and contrast medium load,providing satisfactory image quality for clinical application.
语种中文
原文出处查看原文
文献类型期刊论文
条目标识符http://ir.bjmu.edu.cn/handle/400002259/42958
Collection北京大学第一临床医学院_医学影像科
作者单位北京大学第一医院医学影像科, 北京,100034
Recommended Citation
GB/T 7714
赵永为,郭小超,王霄英,等. 100 kVp条件下碘克沙醇(270 mg I/mL)应用于头颈部CTA 检查的可行性研究[J]. 放射学实践,2014(4):365-368.
APA 赵永为.,郭小超.,王霄英.,雷平贵.,刘建新.,...&高福生.(2014).100 kVp条件下碘克沙醇(270 mg I/mL)应用于头颈部CTA 检查的可行性研究.放射学实践(4),365-368.
MLA 赵永为,et al."100 kVp条件下碘克沙醇(270 mg I/mL)应用于头颈部CTA 检查的可行性研究".放射学实践 .4(2014):365-368.
Files in This Item:
There are no files associated with this item.
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
谷歌学术
谷歌学术Similar articles in
[赵永为]'s Articles
[郭小超]'s Articles
[王霄英]'s Articles
百度学术
百度学术Similar articles in
[赵永为]'s Articles
[郭小超]'s Articles
[王霄英]'s Articles
必应学术
必应学术Similar articles in
[赵永为]'s Articles
[郭小超]'s Articles
[王霄英]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.