Comparison of bioimpedance methods for estimating total body water and intracellular water changes during hemodialysis
Dou, Yanna1,2,3; Liu, Li1,2,3; Cheng, Xuyang1,2,3; Cao, Liyun1,2,3; Zuo, Li1,2,3
关键词Bioimpedance Method Body Fluid Volume Comparison
刊名NEPHROLOGY DIALYSIS TRANSPLANTATION
2011-10-01
DOI10.1093/ndt/gfq854
26期:10页:3319-3324
收录类别SCI
文章类型Article
WOS标题词Science & Technology
类目[WOS]Transplantation ; Urology & Nephrology
研究领域[WOS]Transplantation ; Urology & Nephrology
关键词[WOS]BIOELECTRICAL-IMPEDANCE ANALYSIS ; FLUID ; SPECTROSCOPY ; VOLUMES ; VALIDATION ; ADIPOSITY ; MASS
英文摘要

Background. The accurate assessment of body fluid volume is important in many clinical situations. Hannan et al. proposed a single-frequency bioimpedance equation (HE) to calculate extracellular water (ECW) and total body water (TBW). There are two equations based on the bioimpedance spectroscopy (BIS) method for the evaluation of body fluid volume: Xitron equations (XE) and body composition spectroscopy equations (BCSE). The aim of the study was to compare the accuracy of these three equations in body fluid volume point estimation in maintenance hemodialysis (MHD) patients.

Methods. The BIS method was performed in MHD patients before and after a hemodialysis (HD) session. TBW, ECW and intracellular water (ICW) were calculated by XE, BCSE and HE, respectively. Hydration status (HS) was calculated using inputs of XE, BCSE and HE. ICW before dialysis was compared to ICW after dialysis. The change of TBW and HS using different equations was compared to actual ultrafiltration volume (AUV) that was calculated as weight difference of pre- to postdialysis.

Results. Fifty MHD patients (27 females) were included in the study. Significant changes in ICW were observed using the XE and HE method with ultrafiltration (XE: 15.51 +/- 5.07 versus 16.17 +/- 5.34 L, P < 0.01; HE: 17.40 +/- 5.13 versus 16.55 +/- 4.71 L, P < 0.01). However, no significant ICW change was observed using BCSE (17.47 +/- 4.35 versus 17.54 +/- 4.36 L, P > 0.05). (sic)TBW_XE and (sic)TBW_HE were significantly different from AUV (XE 1.76 +/- 0.89 versus 2.46 +/- 0.89 L, P < 0.01; HE 4.16 +/- 1.36 versus 2.46 +/- 0.89 L, P < 0.01); however, (sic)TBW_BCSE was much closer to AUV (2.27 +/- 0.90 versus 2.46 +/- 0.89 L, P = 0.129). The change of HS using inputs of BCSE was also closer to AUV (2.41 +/- 0.86 versus 2.46 +/- 0.89 L, P = 1.0).

Conclusion. Our study indicated that BCSE provided a better point estimation of ICW and TBW.

语种英语
WOS记录号WOS:000296349200043
项目编号D09050704310902
资助机构Beijing Municipal Science &amp ; Technology Commission
引用统计
被引频次:12[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.bjmu.edu.cn/handle/400002259/50689
专题北京大学第一临床医学院_肾脏内科
作者单位1.Peking Univ, Inst Nephrol, Beijing 100871, Peoples R China
2.Minist Hlth China, Key Lab Renal Dis, Beijing, Peoples R China
3.Peking Univ, Hosp 1, Dept Med, Div Renal, Beijing 100871, Peoples R China
推荐引用方式
GB/T 7714
Dou, Yanna,Liu, Li,Cheng, Xuyang,et al. Comparison of bioimpedance methods for estimating total body water and intracellular water changes during hemodialysis[J]. NEPHROLOGY DIALYSIS TRANSPLANTATION,2011,26(10):3319-3324.
APA Dou, Yanna,Liu, Li,Cheng, Xuyang,Cao, Liyun,&Zuo, Li.(2011).Comparison of bioimpedance methods for estimating total body water and intracellular water changes during hemodialysis.NEPHROLOGY DIALYSIS TRANSPLANTATION,26(10),3319-3324.
MLA Dou, Yanna,et al."Comparison of bioimpedance methods for estimating total body water and intracellular water changes during hemodialysis".NEPHROLOGY DIALYSIS TRANSPLANTATION 26.10(2011):3319-3324.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Dou, Yanna]的文章
[Liu, Li]的文章
[Cheng, Xuyang]的文章
百度学术
百度学术中相似的文章
[Dou, Yanna]的文章
[Liu, Li]的文章
[Cheng, Xuyang]的文章
必应学术
必应学术中相似的文章
[Dou, Yanna]的文章
[Liu, Li]的文章
[Cheng, Xuyang]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。