Performance evaluation of BC-3200 hematology analyzer in a university hospital
Peng, L.; Bai, L.; Nie, L.; Wu, Z.; Yan, C.
关键词Automated Hematology Analyzer Bc-3200 Blood Cell Count Performance
刊名INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY
2008-06-01
DOI10.1111/j.1751-553X.2007.00950.x
30期:3页:205-213
收录类别SCI
文章类型Article
WOS标题词Science & Technology
类目[WOS]Hematology
研究领域[WOS]Hematology
英文摘要

The BC-3200 automated hematology analyzer was evaluated and compared with the Beckman-Coulter AcT (A(c).T diff 2) 3-part differential hematology analyzer. The BC-3200 was evaluated according to guidelines published by the International Committee for Standardization in Hematology (ICSH), Clinical and Laboratory Standards Institute (CLSI), and Department of Food and Drug Administration (FDA). The results demonstrated no background, minimal carryover (< 0.5%), and excellent linearity for hemoglobin (Hb) level, white blood cell (WBC), red blood cell (RBC), and platelet (PLT) counts (> 0.998). Precision was generally acceptable for all complete blood count (CBC) parameters; coefficients of variation (CVs) were within the manufacturer′s claims and CVs of CBC parameters, including WBC, RBC and PLT counts, Hb and mean corpuscular volume, were < 6%. Correlation between the BC-3200 and A(c).T diff 2 was excellent (r > 0.98) for all major CBC parameters (WBC, RBC, and PLT counts and Hb). We conclude that the overall performance of the BC-3200 is excellent and compares well with that of the Coulter A(c).T diff 2.

语种英语
WOS记录号WOS:000255819200006
引用统计
被引频次:2[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.bjmu.edu.cn/handle/400002259/51864
专题北京大学深圳医院
作者单位Peking Univ, Shenzhen Hosp, Dept Lab Med, Shenzhen 518036, Guangdong, Peoples R China
推荐引用方式
GB/T 7714
Peng, L.,Bai, L.,Nie, L.,et al. Performance evaluation of BC-3200 hematology analyzer in a university hospital[J]. INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY,2008,30(3):205-213.
APA Peng, L.,Bai, L.,Nie, L.,Wu, Z.,&Yan, C..(2008).Performance evaluation of BC-3200 hematology analyzer in a university hospital.INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY,30(3),205-213.
MLA Peng, L.,et al."Performance evaluation of BC-3200 hematology analyzer in a university hospital".INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY 30.3(2008):205-213.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Peng, L.]的文章
[Bai, L.]的文章
[Nie, L.]的文章
百度学术
百度学术中相似的文章
[Peng, L.]的文章
[Bai, L.]的文章
[Nie, L.]的文章
必应学术
必应学术中相似的文章
[Peng, L.]的文章
[Bai, L.]的文章
[Nie, L.]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。