IR@PKUHSC  > 北京大学第三临床医学院  > 心血管内科
学科主题临床医学
Usefulness of lumen area parameters determined by intravascular ultrasound to predict functional significance of intermediate coronary artery stenosis
Cui Ming1; Zhu Dan2; Guo Li-jun1,3,4; Song Li-ying2; Zhang Yong-zhen1,3,4; Zhang Fu-chun1,3,4; Niu Jie1,3,4; Wang Gui-song1,3,4; Han Jiang-li1,3,4; Gao Wei1,3,4
关键词Fractional Flow Reserve Intravascular Ultrasound Intermediate Coronary Lesion Minimal Lumen Area
刊名CHINESE MEDICAL JOURNAL
2013-05-05
DOI10.3760/cma.j.issn.0366-6999.20123358
126期:9页:1606-1611
收录类别SCI
文章类型Article
WOS标题词Science & Technology
类目[WOS]Medicine, General & Internal
研究领域[WOS]General & Internal Medicine
关键词[WOS]FRACTIONAL FLOW RESERVE ; INTERVENTION ; SEVERITY ; ANGIOPLASTY ; ANGIOGRAPHY ; CRITERIA ; OUTCOMES ; DISEASE
英文摘要

Background Coronary artery disease is the leading cause of death in China. Percutaneous coronary intervention is a recent milestone technology for treatment coronary artery disease. However, clinical decision making for patients with intermediate coronary stenosis is still controversial. We designed this study to assess the optimal intravascular ultrasound (IVUS) criteria for predicting functional significance of intermediate coronary lesions.

Methods We enrolled 141 patients with 165 intermediate coronary lesions located in vessels with a diameter >= 2.50 mm. IVUS of intermediate coronary lesions were performed before intervention. Pressure-derived fractional flow reserve (FFR) was measured at maximal hyperemia induced by adenosine infusion. An FFR <0.80 was considered as abnormal functional significance.

Results For the overall 165 lesions, the mean FFR value was 0.84 +/- 0.09. The diameter of the stenosis by visual estimation on angiogram was (59.63 +/- 11.29)%. Minimum lumen diameter (MLD), minimum lumen area (MLA) and plaque burden (PB) were (2.00 +/- 0.36) mm, (3.88 +/- 1.34) mm(2), (67.28 +/- 9.89)% respectively by IVUS measurements. An FFR <0.80 was seen in 43 lesions (30.5%). There was a moderate correlation between IVUS parameters and FFR, including MLD (r=0.372, P <0.001), MLA (r=0.442, P <0.001) and PB (r=-0.172, P <0.05). MLA was a predictor for FFR as a continuous variable independent of possible confounding variables (P <0.05), and MLA and PB, were predictors for FFR <0.80 as binary variables (P <0.05). The best cutoff value of MLA to predict FFR <0.80 was <3.15 mm(2), with a 73.6% diagnostic accuracy; sensitivity 71.4%, specificity 67.0%, AUC=0.709, and P <0.001. The cutoff value of the PB to predict FFR <0.80 was 65.45%; sensitivity 82.6%, specificity 41.2%, AUC=0.644, and P <0.01. If both MLA and PB were taken into account, the negative predictive value and the positive predictive value were 88.7% and 64.8% respectively.

Conclusions Anatomic measurements of intermediate coronary lesions obtained by IVUS showed a moderate correlation to FFR values. IVUS-derived MLA >= 3.15 mm(2) may be useful to exclude FFR <0.80, but poor specificity limits its applicability for physiological assessment of lesions <3.15 mm(2). MLA was one of many factors affecting coronary flow hemodynamics. Both MLA and PB should be taken into account when determining functional ischemia.

语种英语
WOS记录号WOS:000319529100002
项目编号Z111107058811003 ; 81070260 ; 81200167 ; 81070260 ; 81200167 ; 7102099
资助机构Capital Characteristic Clinical Application Research from Beijing Science and Technology Committee ; National Natural Science Foundation of China ; Beijing Natural Science Foundation ; Young Core Fund from Peking University Third Hospital
引用统计
被引频次:8[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.bjmu.edu.cn/handle/400002259/52034
专题北京大学第三临床医学院_心血管内科
作者单位1.Peking Univ, Hosp 3, Dept Med, Beijing 100191, Peoples R China
2.Minist Educ, Key Lab Mol Cardiovasc Sci, Beijing 100191, Peoples R China
3.Peking Univ, Hosp 3, Dept Cardiol, Beijing 100191, Peoples R China
4.Minist Hlth, Key Lab Cardiovasc Mol Biol & Regulatory Peptides, Beijing 100191, Peoples R China
推荐引用方式
GB/T 7714
Cui Ming,Zhu Dan,Guo Li-jun,et al. Usefulness of lumen area parameters determined by intravascular ultrasound to predict functional significance of intermediate coronary artery stenosis[J]. CHINESE MEDICAL JOURNAL,2013,126(9):1606-1611.
APA Cui Ming.,Zhu Dan.,Guo Li-jun.,Song Li-ying.,Zhang Yong-zhen.,...&Gao Wei.(2013).Usefulness of lumen area parameters determined by intravascular ultrasound to predict functional significance of intermediate coronary artery stenosis.CHINESE MEDICAL JOURNAL,126(9),1606-1611.
MLA Cui Ming,et al."Usefulness of lumen area parameters determined by intravascular ultrasound to predict functional significance of intermediate coronary artery stenosis".CHINESE MEDICAL JOURNAL 126.9(2013):1606-1611.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Cui Ming]的文章
[Zhu Dan]的文章
[Guo Li-jun]的文章
百度学术
百度学术中相似的文章
[Cui Ming]的文章
[Zhu Dan]的文章
[Guo Li-jun]的文章
必应学术
必应学术中相似的文章
[Cui Ming]的文章
[Zhu Dan]的文章
[Guo Li-jun]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 