北京大学医学部机构知识库
Advanced  
IR@PKUHSC  > 北京大学第三临床医学院  > 期刊论文
学科主题: 临床医学
题名:
Evaluation of the Sysmex UF-1000i Urine Analyzer as a Screening Test to Reduce the Need for Urine Cultures for Urinary Tract Infection
作者: Hu, Xiaozhou; Zhang, Jie; Zhang, Xiaowei
关键词: urinary tract infection ; UF-1000i ; UF-100 ; bacterial count ; WBC count
刊名: LABMEDICINE
发表日期: 2010-06-01
DOI: 10.1309/LM9P7LK3HASE0SKD
卷: 41, 期:6, 页:349-352
收录类别: SCI
文章类型: Article
WOS标题词: Science & Technology
类目[WOS]: Medical Laboratory Technology
研究领域[WOS]: Medical Laboratory Technology
关键词[WOS]: MICROSCOPY ; CYTOMETER ; DIAGNOSIS
英文摘要:

Background: We evaluated the Sysmex UF-1000i flow cytometer as a screening test for urinary tract infections (UTI) and to reduce the need for bacterial cultures.

Methods: By comparing the test results for 308 urine samples with quantitative urine cultures, we established the cutoff criteria for UF-1000i. A bacterial cutoff value of 160/microL provided the best discrimination for UTI, with a sensitivity of 81.1% and a specificity of 83.2% compared with a urine culture. It was possible to forgo 54.9% of cultures with 10% false-negative results.

Results: With a bacterial cutoff value of 10/microL, the sensitivity was improved to 100%, but the specificity declined to 36.8%, and 24.3% of the cultures could be avoided without any false-negative results.

Conclusion: By comparing the test results generated by UF-1000i with those generated by UF-100, we concluded that UF-1000i is better than UF-100 in reducing the need for bacterial cultures. In conclusion, UF-1000i would improve the efficiency of the routine microbiology laboratory and reduce the need for urine cultures for UTI, especially for those induced by bacteria.

语种: 英语
WOS记录号: WOS:000277908300005
Citation statistics:
内容类型: 期刊论文
URI标识: http://ir.bjmu.edu.cn/handle/400002259/55248
Appears in Collections:北京大学第三临床医学院_期刊论文

Files in This Item:

There are no files associated with this item.


作者单位: Peking Univ, Dept Lab Med, Hosp 3, Beijing 100871, Peoples R China

Recommended Citation:
Hu, Xiaozhou,Zhang, Jie,Zhang, Xiaowei. Evaluation of the Sysmex UF-1000i Urine Analyzer as a Screening Test to Reduce the Need for Urine Cultures for Urinary Tract Infection[J]. LABMEDICINE,2010,41(6):349-352.
Service
Recommend this item
Sava as my favorate item
Show this item's statistics
Export Endnote File
Google Scholar
Similar articles in Google Scholar
[Hu, Xiaozhou]'s Articles
[Zhang, Jie]'s Articles
[Zhang, Xiaowei]'s Articles
CSDL cross search
Similar articles in CSDL Cross Search
[Hu, Xiaozhou]‘s Articles
[Zhang, Jie]‘s Articles
[Zhang, Xiaowei]‘s Articles
Related Copyright Policies
Null
Social Bookmarking
Add to CiteULike Add to Connotea Add to Del.icio.us Add to Digg Add to Reddit
所有评论 (0)
暂无评论
 
评注功能仅针对注册用户开放,请您登录
您对该条目有什么异议,请填写以下表单,管理员会尽快联系您。
内 容:
Email:  *
单位:
验证码:   刷新
您在IR的使用过程中有什么好的想法或者建议可以反馈给我们。
标 题:
 *
内 容:
Email:  *
验证码:   刷新

Items in IR are protected by copyright, with all rights reserved, unless otherwise indicated.

 

 

Valid XHTML 1.0!
Copyright © 2007-2017  北京大学医学部 - Feedback
Powered by CSpace