Application of five frozen human-pooled serum samples assigned by the International Federation of Clinical Chemistry and Laboratory Medicine reference procedure in a traceability investigation of gamma-glutamyltransferase catalytic concentration measurements in China
作者: Xia, Changyu1; Tong, Qing2; Wang, Qingtao2; Tang, Zhuling3; Qi, Lili4; Chi, Shan5; Zhang, Man5; Wang, Xuejing1; Li, Haixia1; Xu, Guobin1
刊名: ANNALS OF CLINICAL BIOCHEMISTRY
发表日期: 2010-05-01
DOI: 10.1258/acb.2009.009210
卷: 47, 页:189-194
收录类别: SCI
文章类型: Article
WOS标题词: Science & Technology
类目[WOS]: Medical Laboratory Technology
研究领域[WOS]: Medical Laboratory Technology
关键词[WOS]: FRESH-FROZEN ; CALIBRATION ; HARMONIZATION ; ENZYMES
英文摘要:

Background: The in vitro directive of the European Union requires traceability to the international recommended reference procedures. The application of the reference procedures is necessary in order to evaluate the accuracy of gamma-glutamyltransferase (GGT) assays of routine measurement systems in China.

Methods: Five frozen patient-pooled serum samples were assigned values by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference procedure in order to evaluate the traceability of the results of GGT catalytic activity from six homogeneous systems. One of the serum samples was used to calibrate seven non-homogeneous systems.

Results: All of the homogeneous systems, except the Dade system (Dade Bering Inc, IL, USA), achieved traceability within the measurement range. The Roche and Hitachi systems were better than the other systems. After calibration, the variance of the non-homogeneous systems decreased dramatically from between 14.50% and 25.23% to between 1.25% and 3.09% and the bias decreased from between -11.4% and -4.1% to between 0.5% and 3.5%.

Conclusion: Manufacturers in China should ensure that their calibration systems correspond to the IFCC reference procedures. Fresh frozen pooled patient serum assigned by reference laboratories can be used to calibrate non-homogeneous systems in order to achieve traceability.

语种: 英语
所属项目编号: 2006AA020909
项目资助者: National 863 Plan Projects of China
WOS记录号: WOS:000278696500002
Citation statistics:
内容类型: 期刊论文
URI标识: http://ir.bjmu.edu.cn/handle/400002259/55957
Appears in Collections:北京大学第一临床医学院_期刊论文

Files in This Item:

There are no files associated with this item.


作者单位: 1.Beijing Shijitan Hosp, Clin Lab Dept, Beijing, Peoples R China
2.Biosino Biotechnol & Sci Inc, Beijing, Peoples R China
3.Beijing Chaoyang Hosp, Clin Lab Dept, Beijing, Peoples R China
4.Peking Univ, Clin Lab Dept, Hosp 1, Beijing 100871, Peoples R China
5.Guiyang Med Coll, Clin Lab Dept, Affiliated Hosp, Guizhou, Peoples R China

Recommended Citation:
Xia, Changyu,Tong, Qing,Wang, Qingtao,et al. Application of five frozen human-pooled serum samples assigned by the International Federation of Clinical Chemistry and Laboratory Medicine reference procedure in a traceability investigation of gamma-glutamyltransferase catalytic concentration measurements in China[J]. ANNALS OF CLINICAL BIOCHEMISTRY,2010,47:189-194.
Service
Recommend this item
Sava as my favorate item
Show this item's statistics
Export Endnote File
Google Scholar
Similar articles in Google Scholar
[Xia, Changyu]'s Articles
[Tong, Qing]'s Articles
[Wang, Qingtao]'s Articles
CSDL cross search
Similar articles in CSDL Cross Search
[Xia, Changyu]‘s Articles
[Tong, Qing]‘s Articles
[Wang, Qingtao]‘s Articles
Related Copyright Policies
Null
Social Bookmarking
Add to CiteULike Add to Connotea Add to Del.icio.us Add to Digg Add to Reddit
所有评论 (0)
暂无评论
 
评注功能仅针对注册用户开放,请您登录
您对该条目有什么异议,请填写以下表单,管理员会尽快联系您。
内 容:
Email:  *
单位:
验证码:   刷新
您在IR的使用过程中有什么好的想法或者建议可以反馈给我们。
标 题:
 *
内 容:
Email:  *
验证码:   刷新

Items in IR are protected by copyright, with all rights reserved, unless otherwise indicated.

 

 

Valid XHTML 1.0!
Copyright © 2007-2017  北京大学医学部 - Feedback
Powered by CSpace