Evaluation of analytic and motion-resistant performance of the Mindray 9006 Pulse oximeter
作者: Peng, Liming; Yan, Cunliang; Lu, Hong; Xia, Yong
关键词: noninvasive ; arterial oxygen saturation ; pulse rate ; CO-Oxymetry ; heart rate
刊名: MEDICAL SCIENCE MONITOR
发表日期: 2007-08-01
卷: 13, 期:8, 页:MT19-MT27
收录类别: SCI
文章类型: Article
WOS标题词: Science & Technology
类目[WOS]: Medicine, Research & Experimental
研究领域[WOS]: Research & Experimental Medicine
英文摘要:

Background: The Mindray 9006 is a newly developed pulse oximeter. The performance of noninvasive monitoring ofarterial oxygen saturation by pulse oximetry (SpO(2)) and pulse rate (PR) with the Mindray 9006 was evaluated.

Material/Methods: Twenty healthy volunteers participated in this study. The Mindray 9006 was evaluated according to guidelines including HS3-A, EP5-A, and EP9-A2 published by the CLSI, and ISO 9919. A blinded side-by-side comparison of the Mindray 9006 and the Masimo MS-3 oximeter was also performed.

Results: The results demonstrated that the coefficients of variation (CV) of impression were less than 1% for SpO(2) and less than 5% for PR analyzed by the Mindray 9006. Comparisons of methods showed there was no significant difference in SIO2/PR analysis between the Mindray 9006 with Rapidlab 855 CO-oxymeter or 812 Electrocardiogram (ECG) monitor (p > 0.05) and between the Mindray 9006 and Masimo MS-3 (p > 0.05). There was good correlation for SpO,, between the Mindray 9006 and CO-oxymetry or the Masimo MS-3 (r > 0.96) and for PR between the Mindray 9006 and ECG or the Masimo MS-3 (r > 0.98). The accuracy as the root-mean-square was less dian 3% for both SpO,, and PR between the Mindray 9006 and CO-oxymetry or between the Mindray 9006 and ECG. Also, there was no significant difference for SpO(2)/PR between the motion and rest conditions (p > 0.05).

Conclusions: The overall performance of the Mindray 9006 for SpO(2),, and PR analysis is excellent and compares favorably with the Masimo MS-3.

语种: 英语
WOS记录号: WOS:000248716400014
Citation statistics:
内容类型: 期刊论文
URI标识: http://ir.bjmu.edu.cn/handle/400002259/56151
Appears in Collections:北京大学深圳医院_期刊论文

Files in This Item:

There are no files associated with this item.


作者单位: Peking Univ, Shenzhen Hosp, Dept Lab Med, Shenzhen 518036, Peoples R China

Recommended Citation:
Peng, Liming,Yan, Cunliang,Lu, Hong,et al. Evaluation of analytic and motion-resistant performance of the Mindray 9006 Pulse oximeter[J]. MEDICAL SCIENCE MONITOR,2007,13(8):MT19-MT27.
Service
Recommend this item
Sava as my favorate item
Show this item's statistics
Export Endnote File
Google Scholar
Similar articles in Google Scholar
[Peng, Liming]'s Articles
[Yan, Cunliang]'s Articles
[Lu, Hong]'s Articles
CSDL cross search
Similar articles in CSDL Cross Search
[Peng, Liming]‘s Articles
[Yan, Cunliang]‘s Articles
[Lu, Hong]‘s Articles
Related Copyright Policies
Null
Social Bookmarking
Add to CiteULike Add to Connotea Add to Del.icio.us Add to Digg Add to Reddit
所有评论 (0)
暂无评论
 
评注功能仅针对注册用户开放,请您登录
您对该条目有什么异议,请填写以下表单,管理员会尽快联系您。
内 容:
Email:  *
单位:
验证码:   刷新
您在IR的使用过程中有什么好的想法或者建议可以反馈给我们。
标 题:
 *
内 容:
Email:  *
验证码:   刷新

Items in IR are protected by copyright, with all rights reserved, unless otherwise indicated.

 

 

Valid XHTML 1.0!
Copyright © 2007-2017  北京大学医学部 - Feedback
Powered by CSpace