Evaluation of analytic and motion-resistant performance of the Mindray 9006 Pulse oximeter
Peng, Liming; Yan, Cunliang; Lu, Hong; Xia, Yong
关键词Noninvasive Arterial Oxygen Saturation Pulse Rate Co-oxymetry Heart Rate
刊名MEDICAL SCIENCE MONITOR
2007-08-01
13期:8页:MT19-MT27
收录类别SCI
文章类型Article
WOS标题词Science & Technology
类目[WOS]Medicine, Research & Experimental
研究领域[WOS]Research & Experimental Medicine
英文摘要

Background: The Mindray 9006 is a newly developed pulse oximeter. The performance of noninvasive monitoring ofarterial oxygen saturation by pulse oximetry (SpO(2)) and pulse rate (PR) with the Mindray 9006 was evaluated.

Material/Methods: Twenty healthy volunteers participated in this study. The Mindray 9006 was evaluated according to guidelines including HS3-A, EP5-A, and EP9-A2 published by the CLSI, and ISO 9919. A blinded side-by-side comparison of the Mindray 9006 and the Masimo MS-3 oximeter was also performed.

Results: The results demonstrated that the coefficients of variation (CV) of impression were less than 1% for SpO(2) and less than 5% for PR analyzed by the Mindray 9006. Comparisons of methods showed there was no significant difference in SIO2/PR analysis between the Mindray 9006 with Rapidlab 855 CO-oxymeter or 812 Electrocardiogram (ECG) monitor (p > 0.05) and between the Mindray 9006 and Masimo MS-3 (p > 0.05). There was good correlation for SpO,, between the Mindray 9006 and CO-oxymetry or the Masimo MS-3 (r > 0.96) and for PR between the Mindray 9006 and ECG or the Masimo MS-3 (r > 0.98). The accuracy as the root-mean-square was less dian 3% for both SpO,, and PR between the Mindray 9006 and CO-oxymetry or between the Mindray 9006 and ECG. Also, there was no significant difference for SpO(2)/PR between the motion and rest conditions (p > 0.05).

Conclusions: The overall performance of the Mindray 9006 for SpO(2),, and PR analysis is excellent and compares favorably with the Masimo MS-3.

语种英语
WOS记录号WOS:000248716400014
引用统计
被引频次:1[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.bjmu.edu.cn/handle/400002259/56151
专题北京大学深圳医院
作者单位Peking Univ, Shenzhen Hosp, Dept Lab Med, Shenzhen 518036, Peoples R China
推荐引用方式
GB/T 7714
Peng, Liming,Yan, Cunliang,Lu, Hong,et al. Evaluation of analytic and motion-resistant performance of the Mindray 9006 Pulse oximeter[J]. MEDICAL SCIENCE MONITOR,2007,13(8):MT19-MT27.
APA Peng, Liming,Yan, Cunliang,Lu, Hong,&Xia, Yong.(2007).Evaluation of analytic and motion-resistant performance of the Mindray 9006 Pulse oximeter.MEDICAL SCIENCE MONITOR,13(8),MT19-MT27.
MLA Peng, Liming,et al."Evaluation of analytic and motion-resistant performance of the Mindray 9006 Pulse oximeter".MEDICAL SCIENCE MONITOR 13.8(2007):MT19-MT27.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Peng, Liming]的文章
[Yan, Cunliang]的文章
[Lu, Hong]的文章
百度学术
百度学术中相似的文章
[Peng, Liming]的文章
[Yan, Cunliang]的文章
[Lu, Hong]的文章
必应学术
必应学术中相似的文章
[Peng, Liming]的文章
[Yan, Cunliang]的文章
[Lu, Hong]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。