Evaluation of analytic and motion-resistant performance of the Mindray 9006 Pulse oximeter
作者: Peng, Liming; Yan, Cunliang; Lu, Hong; Xia, Yong
关键词: noninvasive ; arterial oxygen saturation ; pulse rate ; CO-Oxymetry ; heart rate
刊名: MEDICAL SCIENCE MONITOR
发表日期: 2007-08-01
卷: 13, 期:8, 页:MT19-MT27
收录类别: SCI
文章类型: Article
WOS标题词: Science & Technology
类目[WOS]: Medicine, Research & Experimental
研究领域[WOS]: Research & Experimental Medicine
英文摘要:

Background: The Mindray 9006 is a newly developed pulse oximeter. The performance of noninvasive monitoring ofarterial oxygen saturation by pulse oximetry (SpO(2)) and pulse rate (PR) with the Mindray 9006 was evaluated.

Material/Methods: Twenty healthy volunteers participated in this study. The Mindray 9006 was evaluated according to guidelines including HS3-A, EP5-A, and EP9-A2 published by the CLSI, and ISO 9919. A blinded side-by-side comparison of the Mindray 9006 and the Masimo MS-3 oximeter was also performed.

Results: The results demonstrated that the coefficients of variation (CV) of impression were less than 1% for SpO(2) and less than 5% for PR analyzed by the Mindray 9006. Comparisons of methods showed there was no significant difference in SIO2/PR analysis between the Mindray 9006 with Rapidlab 855 CO-oxymeter or 812 Electrocardiogram (ECG) monitor (p > 0.05) and between the Mindray 9006 and Masimo MS-3 (p > 0.05). There was good correlation for SpO,, between the Mindray 9006 and CO-oxymetry or the Masimo MS-3 (r > 0.96) and for PR between the Mindray 9006 and ECG or the Masimo MS-3 (r > 0.98). The accuracy as the root-mean-square was less dian 3% for both SpO,, and PR between the Mindray 9006 and CO-oxymetry or between the Mindray 9006 and ECG. Also, there was no significant difference for SpO(2)/PR between the motion and rest conditions (p > 0.05).

Conclusions: The overall performance of the Mindray 9006 for SpO(2),, and PR analysis is excellent and compares favorably with the Masimo MS-3.

语种: 英语
WOS记录号: WOS:000248716400014
Citation statistics:
内容类型: 期刊论文
URI标识: http://ir.bjmu.edu.cn/handle/400002259/56151
Appears in Collections:北京大学深圳医院_期刊论文

Files in This Item:

There are no files associated with this item.


作者单位: Peking Univ, Shenzhen Hosp, Dept Lab Med, Shenzhen 518036, Peoples R China

Recommended Citation:
Peng, Liming,Yan, Cunliang,Lu, Hong,et al. Evaluation of analytic and motion-resistant performance of the Mindray 9006 Pulse oximeter[J]. MEDICAL SCIENCE MONITOR,2007,13(8):MT19-MT27.
Service
Recommend this item
Sava as my favorate item
Show this item's statistics
Export Endnote File
Google Scholar
Similar articles in Google Scholar
[Peng, Liming]'s Articles
[Yan, Cunliang]'s Articles
[Lu, Hong]'s Articles
CSDL cross search
Similar articles in CSDL Cross Search
[Peng, Liming]‘s Articles
[Yan, Cunliang]‘s Articles
[Lu, Hong]‘s Articles
Related Copyright Policies
Null
Social Bookmarking
Add to CiteULike Add to Connotea Add to Del.icio.us Add to Digg Add to Reddit

Items in IR are protected by copyright, with all rights reserved, unless otherwise indicated.

 

 

Valid XHTML 1.0!
Copyright © 2007-2017  北京大学医学部 - Feedback
Powered by CSpace