Objectives: The aim of this study is to investigate the interference of carbamylated haemoglobin (cHb) to haemoglobin A(1c) (HbA(1c)) measurements in patients with chronic kidney disease (CKD) by three ion-exchange chromatography methods.

Design and methods: We included three ion-exchange chromatography methods (Bio-Rad Variant II, G7 and HA-8160) and one affinity chromatography method (Primus Ultra2) to test HbA(1c). In vitro carbamylation of haemoglobin was also performed to test interference of carbamylation.

Results: 152 CKD patients were diagnosed. The average HbA(1c) and urea levels of nondiabetic and diabetic patients were 5.24% (33.8 mmol/mol) and 7.36% (56.9 mmol/mol) and 16.2 mmol/L and 11.2 mmol/L, respectively. Repeated measures ANOVA showed that the difference of HbA(1c) values between four methods is not statistically significant (P = 0.421 for non-DM and P = 0.336 for DM). With the increase in blood urea concentration from stage 3 to stage 5, there was no corresponding increase in HbA(1c) values for both groups. In vitro carbamylation results showed that carbamylated Hb peaks could be clearly identified by the three ion-exchange methods.

Conclusion: HbA(1c) measurement by these methods may not be impacted by carbamylated Hb. The results should be interpreted with caution in high carbamylated Hb patients. (C) 2013 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

语种英语
WOS记录号WOS:000333763300030
Citation statistics
Cited Times:6[WOS]   [WOS Record]     [Related Records in WOS]
文献类型期刊论文
条目标识符http://ir.bjmu.edu.cn/handle/400002259/58547
Collection北京大学第二临床医学院
作者单位1.Shanghai Ctr Clin Lab, Shanghai 200126, Peoples R China
2.First Hosp Fuzhou, Fuzhou, Fujian Province, Peoples R China
3.Peking Univ, Peoples Hosp, Beijing 100871, Peoples R China
4.First Municipal Peoples Hosp Guangzhou, Guangzhou, Guangdong, Peoples R China
5.Nanjing Med Univ, Nanjing Clin Nucl Med Ctr, Nanjing Hosp 1, Nanjing, Jiangsu, Peoples R China
Recommended Citation
GB/T 7714
Li, Qing,Ju, Yi,Jin, Ting,et al. Haemoglobin A1c measurement in patients with chronic kidney disease[J]. CLINICAL BIOCHEMISTRY,2014,47(6):481-484.
APA Li, Qing.,Ju, Yi.,Jin, Ting.,Pang, Biqiong.,Deng, Jiade.,...&Wang, Hong.(2014).Haemoglobin A1c measurement in patients with chronic kidney disease.CLINICAL BIOCHEMISTRY,47(6),481-484.
MLA Li, Qing,et al."Haemoglobin A1c measurement in patients with chronic kidney disease".CLINICAL BIOCHEMISTRY 47.6(2014):481-484.
Files in This Item:
There are no files associated with this item.
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
谷歌学术
谷歌学术Similar articles in
[Li, Qing]'s Articles
[Ju, Yi]'s Articles
[Jin, Ting]'s Articles
百度学术
百度学术Similar articles in
[Li, Qing]'s Articles
[Ju, Yi]'s Articles
[Jin, Ting]'s Articles
必应学术
必应学术Similar articles in
[Li, Qing]'s Articles
[Ju, Yi]'s Articles
[Jin, Ting]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.