Trueness Assessment for Serum Glucose Measurement Using Commercial Systems through the Preparation of Commutable Reference Materials
作者: Xia, ChangYu1; Liu, Ou2; Wang, LanZhen3; Xu, GuoBing1
关键词: Glucose ; Reference material ; Commercial system ; Bias
刊名: ANNALS OF LABORATORY MEDICINE
发表日期: 2012-07-01
DOI: 10.3343/alm.2012.32.4.243
卷: 32, 期:4, 页:243-249
收录类别: SCI
文章类型: Article
WOS标题词: Science & Technology
类目[WOS]: Medical Laboratory Technology
研究领域[WOS]: Medical Laboratory Technology
关键词[WOS]: GENERAL CLINICAL-CHEMISTRY ; PROJECT
英文摘要:

Background: Commutable reference materials (RMs) are suitable for end-users for evaluating the metrological traceability of values obtained using routine measurement systems. We assessed the performance of 6 routine measurement systems with validated secondary RMs.

Methods: We tested the homogeneity, stability, and commutability of 5 minimally processed human serum pools according to the standard guidelines. The serum pools were assigned values as per the reference procedure of the United States Centers for Disease Control and were used to evaluate the trueness of results from 6 commercial measurement systems based on enzymatic methods: 3 glucose oxidase (GOD) and 3 hexokinase (HK) methods.

Results: The prepared RMs were validated to be sufficiently homogenous, stable, and commutable with the patient samples. Method bias varied for different systems: GOD01, -0.17 to 2.88%; GOD02, 1.66 to 4.58%; GOD03, -0.17 to 3.14%; HK01, -3.48 to -0.85%; HK02, -3.83 to -0.11%, and HK03, -1.82 to -0.27%.

Conclusions: We observed that the prepared serum glucose RMs were qualified for trueness assessment. Most of the measurement systems met the minimal quality specifications.

语种: 英语
所属项目编号: 2006AA020909
项目资助者: National 863 Plan Projects ; Capital Foundation of Medical Development of China
WOS记录号: WOS:000306207400001
Citation statistics:
内容类型: 期刊论文
URI标识: http://ir.bjmu.edu.cn/handle/400002259/61690
Appears in Collections:北京大学第一临床医学院_期刊论文

Files in This Item:

There are no files associated with this item.


作者单位: 1.Beijing Leadman Biochem Co Ltd, Reference Lab, Beijing, Peoples R China
2.Peking Univ, Hosp 1, Dept Clin Lab, Beijing 100034, Peoples R China
3.Tsinghua Univ, Hosp 1, Dept Clin Lab, Beijing 100084, Peoples R China

Recommended Citation:
Xia, ChangYu,Liu, Ou,Wang, LanZhen,et al. Trueness Assessment for Serum Glucose Measurement Using Commercial Systems through the Preparation of Commutable Reference Materials[J]. ANNALS OF LABORATORY MEDICINE,2012,32(4):243-249.
Service
Recommend this item
Sava as my favorate item
Show this item's statistics
Export Endnote File
Google Scholar
Similar articles in Google Scholar
[Xia, ChangYu]'s Articles
[Liu, Ou]'s Articles
[Wang, LanZhen]'s Articles
CSDL cross search
Similar articles in CSDL Cross Search
[Xia, ChangYu]‘s Articles
[Liu, Ou]‘s Articles
[Wang, LanZhen]‘s Articles
Related Copyright Policies
Null
Social Bookmarking
Add to CiteULike Add to Connotea Add to Del.icio.us Add to Digg Add to Reddit

Items in IR are protected by copyright, with all rights reserved, unless otherwise indicated.

 

 

Valid XHTML 1.0!
Copyright © 2007-2017  北京大学医学部 - Feedback
Powered by CSpace