Stapes prosthesis attachment: The effect of crimping on sound transfer in otosclerosis surgery
Huber, AM; Ma, FR; Felix, H; Linder, T
关键词stapes surgery stapedotomy otosclerosis prosthesis laser Doppler
刊名LARYNGOSCOPE
2003-05-01
113期:5页:853-858
收录类别SCI
文章类型Article
WOS标题词Science & Technology
类目[WOS]Medicine, Research & Experimental ; Otorhinolaryngology
研究领域[WOS]Research & Experimental Medicine ; Otorhinolaryngology
关键词[WOS]STAPEDECTOMY
英文摘要

Objectives/Hypothesis: Although in stapes surgery successful hearing improvement may be achieved in the majority of patients, unsatisfactory closure of the air-bone gap can be recorded. One of many reasons for unexpected failures of stapes surgery may be the insufficient crimping of a stapes prosthesis onto the incus. The objectives of the study were to assess the amount of sound transmission loss in response to the quality of prosthesis crimping and to identify a required loop attachment pattern to obtain good sound transmission results. Study Design: Experimental. Methods A temporal bone model was developed to measure the sound transmission properties between incus and prosthesis on 17 fresh human temporal bones. The attachment of a titanium stapes piston was assessed without crimping, followed by loose crimping and tight fixation to the incus, using scanning laser Doppler interferometry, endoscopic photography, micro grinding technique, and scanning electron microscopy. An algorithm had to be developed to simulate acoustical stimulation using electromechanical stimulation. Results: Optimal tight crimping of the stapes piston revealed consistent good sound transfer function ranging from 0 to 7 dB loss, and loss remained, on average, at 2 dB. The mean transmission losses for conditions of loose crimping and no crimping were surprisingly small (within 10 dB). However, these unusual crimping conditions allowed a wide range of losses up to 28 dB. A close coupling at least at two opposite points was obligatory to obtain consistently good results. Conclusions: Perfect hearing reconstruction necessitates ideal crimping of a prosthesis to obtain consistently good results. However, the final functional gain depends on many different intraoperative and postoperative factors.

语种英语
WOS记录号WOS:000182729200015
引用统计
被引频次:53[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.bjmu.edu.cn/handle/400002259/64307
专题北京大学第三临床医学院_耳鼻喉科
作者单位1.Univ Zurich Hosp, Dept Otorhinolaryngol Head & Neck Surg, CH-8091 Zurich, Switzerland
2.Beijing Med Univ, Third Teaching Hosp & Clin Med Coll, Dept Otolaryngol, Beijing 100083, Peoples R China
3.Kantonsspit Luzern, Dept Otorhinolaryngol Head & Neck Surg, Luzern, Switzerland
推荐引用方式
GB/T 7714
Huber, AM,Ma, FR,Felix, H,et al. Stapes prosthesis attachment: The effect of crimping on sound transfer in otosclerosis surgery[J]. LARYNGOSCOPE,2003,113(5):853-858.
APA Huber, AM,Ma, FR,Felix, H,&Linder, T.(2003).Stapes prosthesis attachment: The effect of crimping on sound transfer in otosclerosis surgery.LARYNGOSCOPE,113(5),853-858.
MLA Huber, AM,et al."Stapes prosthesis attachment: The effect of crimping on sound transfer in otosclerosis surgery".LARYNGOSCOPE 113.5(2003):853-858.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Huber, AM]的文章
[Ma, FR]的文章
[Felix, H]的文章
百度学术
百度学术中相似的文章
[Huber, AM]的文章
[Ma, FR]的文章
[Felix, H]的文章
必应学术
必应学术中相似的文章
[Huber, AM]的文章
[Ma, FR]的文章
[Felix, H]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。