Hemodynamic study of overlapping bare-metal stents intervention to aortic aneurysm
Zhang, Peng1; Sun, Anqiang1; Zhan, Fan1; Luan, Jingyuan2; Deng, Xiaoyan1
关键词Thoracic aortic aneurysm Overlapping bare-metal stents Hemodynamics Numerical simulation
刊名JOURNAL OF BIOMECHANICS
2014-11-07
DOI10.1016/j.jbiomech.2014.08.025
47期:14页:3524-3530
收录类别SCI
文章类型Article
WOS标题词Science & Technology
类目[WOS]Biophysics ; Engineering, Biomedical
研究领域[WOS]Biophysics ; Engineering
关键词[WOS]ENDOVASCULAR REPAIR ; MULTILAYER STENT ; THORACOABDOMINAL ANEURYSMS ; INTERACTION SIMULATION ; ARTERY ANEURYSM ; FLOW PATTERNS ; ENDOLEAKS ; PERFORMANCE ; OUTCOMES ; STRESS
英文摘要

To investigate the hemodynamic performance of overlapping bare-metal stents intervention treatment to thoracic aortic aneurysms (TAA), three simplified TAA models, representing, no stent, with a single stent and 2 overlapped stents deployed in the aneurismal sac, were studied and compared in terms of flow velocity, wall shear stress (WSS) and pressure distributions by means of computational fluid dynamics. The results showed that overlapping stents intervention induced a flow field of slow velocity near the aneurismal wall. Single stent deployment in the sac reduced the jet-like flow formed prior to the proximal neck of the aneurysm, which impinged on the internal wall of the aneurysm. This jet-like flow vanished completely in the overlapping double stents case. Overlapping stents intervention led to an evident decrease in WSS; meanwhile, the pressure acting on the wall of the aneurysm was reduced slightly and presented more uniform distribution. The results therefore indicated that overlapping stents intervention may effectively isolate the thoracic aortic aneurysm, protecting it from rupture. In conclusion, overlapping bare-metal stents may serve a purpose similar to that of the multilayer aneurysm repair system (MARS) manufactured by Cardiatis SA (Isnes, Belgium). (C) 2014 Elsevier Ltd. All rights reserved.

语种英语
WOS记录号WOS:000345183500014
项目编号11332003 ; 11421202 ; 11228205 ; 11202016 ; 11102014 ; 2012BAI18B02 ; 2012BAI14B04
资助机构National Natural Science Foundation of China ; National Key Technology RD Program
引用统计
被引频次:10[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.bjmu.edu.cn/handle/400002259/64365
专题北京大学第三临床医学院
作者单位1.Beihang Univ, Sch Biol Sci & Med Engn, Minist Educ, Key Lab Biomech & Mechanobiol, Beijing 100191, Peoples R China
2.Peking Univ, Dept Intervent Radiol & Vasc Surg, Hosp 3, Beijing 100191, Peoples R China
推荐引用方式
GB/T 7714
Zhang, Peng,Sun, Anqiang,Zhan, Fan,et al. Hemodynamic study of overlapping bare-metal stents intervention to aortic aneurysm[J]. JOURNAL OF BIOMECHANICS,2014,47(14):3524-3530.
APA Zhang, Peng,Sun, Anqiang,Zhan, Fan,Luan, Jingyuan,&Deng, Xiaoyan.(2014).Hemodynamic study of overlapping bare-metal stents intervention to aortic aneurysm.JOURNAL OF BIOMECHANICS,47(14),3524-3530.
MLA Zhang, Peng,et al."Hemodynamic study of overlapping bare-metal stents intervention to aortic aneurysm".JOURNAL OF BIOMECHANICS 47.14(2014):3524-3530.
条目包含的文件
条目无相关文件。
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Zhang, Peng]的文章
[Sun, Anqiang]的文章
[Zhan, Fan]的文章
百度学术
百度学术中相似的文章
[Zhang, Peng]的文章
[Sun, Anqiang]的文章
[Zhan, Fan]的文章
必应学术
必应学术中相似的文章
[Zhang, Peng]的文章
[Sun, Anqiang]的文章
[Zhan, Fan]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。